Function£ºPut the spliced crystal bar and flitch together on YX-6F orientation instrument, let the bottom of flitch rely on base surface of regular instrument, open the optical shutter to do rechecking measurement. YX-6F can do recheck measurement of X axis for spliced crystal bar, the result can be compared with print result by YX-6D, if happen difference, can remove the crystal bar from flitch and splice again, make sure splice accuracy of crystal bar before cutting.
|
X-ray tube |
copper target with anode grounding, force air cooling |
|
Tube voltage |
30KVP£¬switch closed under full voltage |
|
Tube current |
0---5mA£¬continuous adjustable. |
|
Input power |
Not over 0.3KW |
|
Orientation accuracy |
¡À30¡å |
|
Worm gear running error |
Not over 30¡å |
|
Counter rotation angle |
-10¡ã------+100¡ã |
|
Sample rotation angle |
-10¡ã------+50¡ã |
|
Counter tube |
Geiger tube, voltage of 600-----1100V |
|
Time constant |
0.1¡¢0.4¡¢3second; 3levels adjustable |
|
Protection |
High voltage interlock protection, temperature protection |
|
slit |
4¡ä¡¢5¡ä¡¢6¡ä |
|
Outer dimension |
1200mm¡Á650mm¡Á1100mm |
|
Equipment weight |
250KG |
|
Sample diameter |
Silicon single crystal ¦µ3---¦µ8inch silicon wafer£» |
|
Measured crystal bar length |
Silicon single crystal ¦µ3¡ª8inch£»500mm length crystal bar |
|